Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Class H insulation film dielectric strength testing is a critical quality control and qualification procedure for heat-resistant insulating films rated for continuous operation at 180°C. This technical guide explains the applicable standards, sample requirements, electrode configurations, and voltage application methods used to determine the breakdown voltage and dielectric strength of polyimide and other Class H films used in motors, transformers, and aerospace systems.

Class H Insulation Film Dielectric Strength Testing

Technical Overview

Class H insulation, classified under IEC 60085 as a thermal class with a maximum continuous operating temperature of 180°C, represents one of the highest thermal endurance ratings for organic insulating materials. Films in this category, most notably polyimide (PI) films and polyimide-based composites, are widely used in rotating machines, transformers, and aerospace electrical systems where sustained exposure to elevated temperatures and strong electric fields occurs simultaneously.

Dielectric strength testing measures the ability of an insulation film to withstand an applied voltage without experiencing dielectric breakdown, which is the catastrophic loss of insulating properties resulting in conductive failure through the material. The result is expressed either as breakdown voltage in kilovolts (kV) or, more commonly for material comparison, as dielectric strength in kilovolts per millimeter (kV/mm). For thin polyimide films in the 25 μm to 125 μm range, typical short-time breakdown strength values fall between approximately 100 kV/mm and 300 kV/mm, with thinner films generally exhibiting higher intrinsic strength values per unit thickness.

Because dielectric performance directly determines the reliability, creepage allowances, and safety margins of electrical insulation systems, accurate and repeatable breakdown testing is essential for material qualification, incoming inspection, and design verification of Class H film products.

Testing Samples

Polyimide Base Films

Standard specimens consist of polyimide film in common commercial thicknesses, including 25 μm, 50 μm, 75 μm, and 125 μm. Specimens should be visually inspected before testing and rejected if they contain pinholes, wrinkles, contamination, or edge damage, since such defects can cause premature breakdown that does not reflect the true material capability.

Composite and Specialty Films

In addition to homogeneous polyimide film, testing also covers Class H composite products such as polyimide films laminated with fluoropolymer (PI/FEP) layers for heat-sealing applications, corona-resistant polyimide films incorporating inorganic nanofillers for inverter-fed motor duty, and mica-polyimide flexible tapes used in high-voltage rotating machine insulation. Each layer structure must be tested in its as-delivered condition unless otherwise specified.

Sample Preparation and Conditioning

Specimens are typically cut to dimensions of at least 100 mm by 100 mm to guarantee adequate creepage distance beyond the electrode perimeter. Unless otherwise specified by the product standard, samples are conditioned in a controlled atmosphere of 23 ± 2°C and 50 ± 5% relative humidity for a minimum of 24 hours prior to testing. Thickness is measured at multiple points within the active test area, and the mean value is used for the dielectric strength calculation.

Testing Items

Depending on the purpose of the evaluation, the following electrical characteristics are typically assessed on Class H insulation films:

  • AC dielectric breakdown voltage at power frequency (48–62 Hz), expressed in kV rms
  • DC dielectric breakdown voltage for DC transmission and converter applications
  • Dielectric strength in kV/mm, calculated as breakdown voltage divided by specimen thickness
  • Elevated-temperature breakdown performance at 180°C to verify behavior at the Class H thermal limit
  • Statistical breakdown distribution parameters, including Weibull scale and shape values for reliability assessment
  • Complementary indicators such as partial discharge inception voltage and corona resistance life when required by the insulation system specification

Testing Methods

Applicable Standards

The primary international and national standards governing this test include the following:

  • IEC 60243-1, covering electric strength tests of solid insulating materials at power frequency
  • ASTM D149, defining dielectric breakdown voltage and dielectric strength measurement methods
  • GB/T 1408.1, the Chinese national method widely referenced for domestic product certification
  • IEC 60626 series, applicable to composite flexible materials for electrical insulation

Electrode Configuration

The most common arrangement for film testing uses two symmetric metal cylinder electrodes of 25 mm diameter with rounded edges, pressed against opposite faces of the specimen under a defined contact force. Alternative configurations permitted by the standards include sphere-to-plane and sphere-to-sphere electrode systems, which are selected when the referenced material specification requires them. Electrode surfaces must be clean, polished, and free of pitting or carbon deposits to avoid localized field concentration.

Voltage Application Procedures

The standards define several recognized voltage application sequences, and the chosen method must be stated in the test report:

  • Short-time method: voltage is raised uniformly from zero at a constant rate, commonly 500 V/s, until breakdown occurs
  • Slow rate-of-rise method: a lower ramp rate is applied when a more gradual stress increase is representative of service conditions
  • Step-by-step method: voltage starts at a predetermined fraction of the expected breakdown value and increases in defined fixed increments at fixed time intervals until failure

Test Medium and Environment

For thin films with high intrinsic strength, testing in air can trigger surface flashover before true breakdown. Immersion of the electrode assembly in a clean insulating liquid, typically filtered transformer oil, suppresses flashover and permits accurate measurement of through-thickness breakdown. Tests may also be performed inside an environmental chamber when breakdown at elevated temperature is required.

Data Evaluation and Reporting

At least five valid breakdowns are normally required per condition, with flashover events and edge failures excluded and retested. The dielectric strength is calculated by dividing each breakdown voltage by the corresponding thickness, and the report includes the individual values, mean, standard deviation, and minimum result. For design-oriented reliability evaluation, Weibull statistical analysis is applied to derive the characteristic breakdown strength and scatter parameters.

Applications

Verified dielectric strength data for Class H insulation films underpins the design and certification of numerous high-reliability electrical products:

  • Traction and crane motors, metallurgical duty motors, and other rotating machines with Class H thermal classification
  • Dry-type power and special transformers operating at elevated temperatures
  • Electric vehicle drive motors and inverter-fed systems requiring corona-resistant film insulation
  • Aerospace wiring, flexible printed circuits, and avionics insulation operating across wide temperature ranges
  • Generator slot insulation, magnet wire wraps, and phase barrier layers in high-voltage rotating machines
  • Magnet and coil insulation in wind turbine generators and industrial heating equipment

In each of these applications, the measured breakdown margin, often evaluated against the applicable safety factor, ensures that the film can endure operating voltages, transient overvoltages, and thermal aging without insulation failure throughout the expected service life.

FAQ

What does Class H mean in insulation grading?

Class H designates an insulating material or system with a maximum continuous operating temperature of 180°C as defined by IEC 60085. Materials in this class retain adequate electrical and mechanical properties after prolonged thermal exposure at this temperature.

Which test standard should be selected for Class H film?

IEC 60243-1 and ASTM D149 are the most widely accepted methods for dielectric strength determination. The choice depends on the customer specification, target market, and any product standard that mandates a particular electrode system or voltage application sequence.

Why is insulating oil used during the test?

Thin, high-strength films can flash over across the air gap around the electrodes before internal breakdown occurs. Immersion in insulating oil prevents this surface discharge and allows the genuine through-thickness breakdown voltage of the film to be measured.

How does temperature affect dielectric strength results?

Breakdown strength generally decreases as temperature rises toward the thermal class limit. Testing at 180°C is therefore recommended when the film will operate continuously at the Class H temperature, since room-temperature data alone may overstate the available insulation margin.

How many specimens should be tested?

A minimum of five valid breakdowns per condition is standard practice, while ten or more specimens are preferred when Weibull statistical analysis is required for reliability characterization or design input.

What factors most commonly cause low or scattered results?

Defects such as pinholes and inclusions, incorrect thickness measurement, contaminated or damaged electrode surfaces, excessive humidity during conditioning, and surface flashover events are the leading causes of anomalous breakdown data and should be systematically controlled.